Einleitung
This guide details how to remove the Back Panel of the Fluke 77 Series III Multimeter. The Back Panel Replacement guide is a prerequisite-only guide for LCD and PCA Board replacement guides.
Werkzeuge
-
-
Turn the multimeter so it is display side down.
-
Lift up the back stand of the device by putting your thumb in the curved well at the bottom of the device, so you have access to the two covered screws below.
-
-
-
Turn the Fluke 77 Series III Multimeter back over so that it is display side up.
-
Pry the front panel from the back panel at the seam.
-
If the front panel does not come off easily, move a plastic opening tool along the seam until the panels separate completely.
-
-
-
Locate the 9V battery in the lower right corner of the device.
-
Pull the battery out of its terminal.
-
-
-
Locate the fuses near the bottom left corner inside the device.
-
The .440 Amp fuse is located in the upper left hand fuse holder.
-
The 11 Amp fuse is located in the lower right hand fuse holder.
-
Lift the rounded edge of the fuse with one finger until that side is completely out of its holder.
-
-
-
Grab the lifted edge of the fuse between your thumb and forefinger and pull it the rest of the way out of the device.
-
Repeat steps 4 and 5 to remove the other fuse if you are replacing both fuses or if you are replacing a different part further on.
-
-
-
Locate the rubber range and hold buttons in the internal assembly. These are found just below the LCD screen on the left hand side.
-
Remove the rubber range and hold button part with your thumb and forefinger.
-
-
-
Locate the silver 7 mm screw found underneath the 11A fuse.
-
Using the Phillips #0 Screwdriver, unscrew and remove the silver 7 mm Phillips #0 screw that attaches the back panel to the internal cover.
-
To reassemble your device, follow these instructions in reverse order.