If you need to replace the air and oxygen flow sensors on the newport e500 you will need to start with the prerequisites for this operation. You will need a medium and small sized Philips screwdriver and a torque wrench. For more information refer to page 59 of the service manual.
Note: When reinstalling flow sensors, note whether the sensor is labeled for air or oxygen and be sure to install in the appropriate position.
Werkzeuge
-
-
Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
-
-
-
Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
-
-
-
Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).
-
-
-
Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.
-
To reassemble your device, follow these instructions in reverse order.