Einleitung
This guide will show you how to remove the pneumatic panel for further disassembly and replacement. For some helpful tips, go to page 62 in the user manual.
Werkzeuge
-
-
Remove the four hex nuts and lock washers (two at each hinge) that secure the cover to the hinges on the left side of the pneumatic box.
-
-
-
Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
-
-
-
Gently pull each flow sensor straight out from the servo valve, (the flow sensor push-fits into the servo valve port).
-
-
-
Disconnect the flow sensor cable from its 10 pin connector at the underside of the flow sensor PCB and lift out the flow sensor.
-
-
-
Remove the screw and washer that secure the exhalation manifold support to the pneumatic panel.
-
-
-
Remove the standoff and five screws and washers that secure the pneumatic panel to the pneumatic box.
-
-
-
Lift the pneumatic panel out of the box, and from the underside, detach all tubes and the heater wires that connect to the analog PCB.
-
To reassemble your device, follow these instructions in reverse order.