Einleitung
The air inlet is located on the right side panel of the device. According to the service manual, the air inlet filter should be inspected and replaced as required. Environments with more lint and dust may require more frequent replacement.
PLEASE NOTE: This repair guide was developed by the iFixit team based on Philips's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Philips.
Replacing the air inlet filter is discussed in:
- chapter 9 “Care and Maintenance” of user manual
- chapter 7.2 “Component Removal/Installation - Air Inlet Filter” of service manual
-
-
Press and release the ON/Shutdown key, opening the Shutdown window in the screen.
-
Select Ventilator Shutdown, which will shut down the ventilator.
-
-
-
Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.
-
-
-
Remove the inlet filter by pinching it out of the recess in the bracket.
-
Install a new air filter by tucking it into the recessed area.
-
- Place back the side panel.
- Connect the ventilator back to the AC power.
- Power on the ventilator with the ON/Shutdown key.
- Run the preoperational check.