The Pre-Use Check includes tests and measurements of:
- internal technical functionality
- internal leakage
- pressure transducers
- O2 cell / O2 sensor
- flow transducers
- safety valve
- battery modules
- patient breathing system leakage
- circuit compliance
WARNINGS
- Always perform a Pre-Use Check before connecting the ventilator to a patient.
- The separate Patient Circuit Test that can be performed in Standby mode does not replace the Pre-Use Check.
- Do not connect the ventilator to a patient while a malfunction persists.
- Do not lift the expiratory cassette while the ventilator is in operation; instead, lift the cassette while in Standby mode.
Important: If you change the breathing circuit after calculating the circuit compliance compensation factor, perform a new Pre-Use Check.