Einleitung

The Pre-Use Check includes tests and measurements of:

  • internal technical functionality
  • internal leakage
  • pressure transducers
  • O2 cell / O2 sensor
  • flow transducers
  • safety valve
  • battery modules
  • patient breathing system leakage
  • circuit compliance

WARNINGS

  • Always perform a Pre-Use Check before connecting the ventilator to a patient.
  • The separate Patient Circuit Test that can be performed in Standby mode does not replace the Pre-Use Check.
  • Do not connect the ventilator to a patient while a malfunction persists.
  • Do not lift the expiratory cassette while the ventilator is in operation; instead, lift the cassette while in Standby mode.

Important: If you change the breathing circuit after calculating the circuit compliance compensation factor, perform a new Pre-Use Check.