My device name is GE Tec 6, it is the 6th model of the GE Tec series going up to 7 and starting from 5. It was developed in 1998. Some distinguishing features of the GE Tec 6 Plus are that the Tec 6 Plus consist of a large, 425 mL sump capacity to accommodate nearly two full bottles of Desflurane, letting you use the vaporizer for extended periods before refilling, refilling can be done while the vaporizer is in use, the exclusive Saf-T-Fill valve of the Desflurane bottle is designed to fit only the complementary filling device of the Tec 6 Plus, so you can refill quickly. This valve helps prevent accidental release of vapour into the OR and helps to ensure that only Desflurane is introduced into the Tec 6 Plus. Some major facts are that:

  • Voltage requirement: 100 to 120V AC 50/60 Hz | 220 to 240V AC 50/60 Hz
  • Concentration range: 1 to 18%
  • Anesthetic Agent: Des / Desflurane

Technical Specifications

Dimensions

Weight: 21 lbs (9.5 kg)                Depth: 9.25 in (235 mm)

Width: 4.33 in (110 kg)                Height: 9.84 in (250 mm)

Performance

  • Scale: 1% from 1 to 10% | 2% from 10 to 18%
  • Calibrated flow range at 21 ± 2° C: 0.2 to 10 liters/minute
  • Accuracy at 5 liters/min O2: ± 0.5% v/v of delivered agent or ± 15% dial setting (whichever is greater)
  • Liquid capacity: 390 ml nominal volume (indicated on level display) | 240 ml refill capacity indicator nominal ml reserve (not indicated on level display)
  • Flow resistance: Less than 50 cm H2O at 5 liters/minute of O2 at 21 °C | Less than 250 cm H2O at 15 liters/minute of O2 at 21 °C
  • Battery: Duracell MN 1604 or VARTA Energy 2000
  • Operating temperature range: 18 °C to 30 °C (64 °F to 86 °F)
  • Operating humidity range: 30% to 75% relative humidity (non-condensing)
  • Storage temperature range: -40 °C to 60 °C (-40 °F to 140°F)
  • Storage relative humidity range: 0% to 95% relative humidity at 40 °C (non-condensing)
  • Storage atmospheric pressure: 70.0 to 106.0 kPa