Background and Identification

The Mallinckrodt Nellcor NPB-75 is a handheld capnograph and pulse oximeter sold by Mallinckrodt Pharmaceuticals. Pulse oximeters are non-invasive devices used to monitor a patient’s oxygen saturation. Mallinckrodt is a manufacturer of specialty pharmaceuticals, imaging agents, and generic drugs. The company also acquires, manufactures, and sells products for diagnostic procedures and pain treatment.

The NPB-75 model is designed to continuously monitor end-tidal carbon dioxide levels, respiratory rate, oxygen saturation levels, and pulse rate. It can be used on adult, pediatric, and infant/neonatal patients. The device is intended for continuous, noninvasive monitoring for hospital and mobile use.

This pulse oximeter can be powered either by AC power or on its internal Nickel Metal Hydride battery pack, which is rechargeable. The NPB-75 model can be operated using a four-key keypad. End-tidal carbon dioxide readings are displayed on the left-most 7-segment LED digital display. Oxygen saturation readings are shown on the right-most LED digital display. Other information like alarms and waveforms can be displayed on the device’s LCD graphic display, which is the large display near the center of the device’s front face.

The Mallinckrodt Nellcor NPB-75 is sometimes sold under the Puritan Bennett brand name. For these models, the name “NELLCOR PURITAN BENNETT” is printed near the bottom of the device’s front face. The model name, “NPB-75” is printed in the upper left-hand corner of the monitor’s front face.

Technical Specifications

PHYSICAL:

  • Size: 206 mm H x 88 mm W x 52.5 mm D (8.11”H x 3.46” W x 2.06”D)
  • Weight: 850 grams (1.87 lb.) (including battery pack)
  • Noise Emission: maximum 45 dB(A)

ENVIRONMENTAL:

  • Temperature
    • Operating: 0 oC to 45 oC (32 oF to 113 oF)
    • Storage: -35 oC to 70 oC (-31 oF to 158 oF)
  • Relative Humidity: 10 to 95% (noncondensing)
  • Pressure and Altitude (for operating and storage):
    • Pressure: 430 mmHg to 795 mmHg
    • Altitude: -380m to 4,570m (-1,250 ft. to 15,000 ft.)

SAFETY STANDARDS: The monitor was designed to comply with EN60601-1/1990, A1/1993, A2/1995, UL2601-1, CSA C22.2 No. 601.1-M90, ISO 10993-1 (Sensors)

PERFORMANCE:

Capnograph

Sampling Rate: 50 ±7.5 ml/min.

CO2 Range: 0-99 mmHg (0-13.2 kPa and 0-13.0 Vol%) at sea level

Accuracy:

  • The CO2 reading reaches its steady-state accuracy 20 minutes after power-up.
    • 0 - 38 mmHg: (±2 mmHg)
    • 39 - 76 mmHg: (±5% of reading)
    • 77 - 99 mmHg: (±8% of reading)
  • From 3 minutes after power-up until steady state is reached, the CO2reading accuracy is:
    • 0 - 38 mmHg: (±4 mmHg)
    • 39 - 76 mmHg: (±12% of reading)
    • 77 - 99 mmHg: (±12% of reading)

Equivalent values for kPa and Vol%

Frequency Response: EtCO2 accuracy is maintained up to 80 breaths/min. (for maintaining accuracy for respiration rate over 60 bpm, use the neonatal mode.) From 81 to 150 bpm accuracy is ±12%, if the EtCO2 is higher than 18.8 mmHg in neonatal mode.

Respiration Rate: 0-150 breaths/min.

Warm-up Time: 30 seconds (typical)

Rise Time: 190 msec for neonatal mode with low dead space endotracheal tube adapter; 240 msec for adult mode with Microstream Airway Adapter

System Response Time: 2.45 seconds (typical), 2.9 seconds maximum (includes delay time and rise time)

Ambient Pressure: Compensated internally - automatic

Alarms: EtCO2 high, EtCO2 low, FiCO2 high, Apnea message.

Pulse Oximeter

Display Ranges: SpO2 Saturation Range: 0-100% SpO2

Saturation Accuracy: (% SpO2, ±1 SD)

  • Adults: 70-100%±2 digits
    • 0 - 69% Unspecified
  • Neonates 70-100%±3 digits
    • 0 - 69% Unspecified

Pulse Rate Range: 20-250 beats per minute (bpm)

Pulse Rate Accuracy: ± 3 bpm

Alarms: SpO2 high, SpO2 low, Pulse high, Pulse low

POWER SPECIFICATIONS:

External Power Source: 12V DC Medical Grade Adapter

Internal Power Source: Ni-MH Rechargeable Battery Pack 7.2V 2100 mA/h (intended for continuous operation)

Operating Time (fully charged): Between 3 to 6 hours depending on power management. This spec reflects the performance of a new battery; age and usage will degrade capacity.

Recharging Period: Approximately 4.5 hours internal recharging

Charger Type: Internal

COMPONENTS AND USER INTERFACE:

Displays:

  • LCD graphic display (128 x 64 dots) with LED backlight dimension 75 mm x 53 mm.
  • Two numeric fields 3 digits each, using 7-segment LED graphic display dimension 22mm x 14 mm.
  • Two-color alarm bar yellow, red

Controls and Indicators:

  • Front Panel: ON/OFF switch; Alarm Silence/Alarm Menu button; Contrast/Value Change button; Event/Home button; Next/Menu button.

Connections:

  • Front Panel: CO2 Input connector
  • Top Panel: SpO2 Input socket
  • Rear Panel: Clamp connector, battery release button
  • Side Panel: Power Supply/Communication Adapter Port, Gas output connector

Additional Information

Wikipedia: Mallinckrodt

Mallinckrodt Nellcor NPB-75 Service Manual

Mallinckrodt Nellcor NPB-75 User Guide