The Nellcor Bedside Capnograph Pulse Oximeter is a pulse oximeter made by Medtronic's Nellcor division. It is sometimes referred to by its model number, PM100N. Pulse oximeters are non-invasive devices used to monitor a patient’s oxygen saturation. The Nellcor division focuses mainly on producing pulse oximeters and oxygen monitors.
The PM100N model is a handheld pulse oximeter and capnograph used to monitor respiratory rate, end-tidal carbon dioxide, pulse rate, and oxygen saturation. This device is suitable for continuously monitoring adult, pediatric, and neonatal patients. It can be used for hospital and mobile use, as long as the machine is protected from excessive moisture. The monitor is powered either by AC power or by a rechargeable Nickel Metal Hydride battery pack.
The PM100N pulse oximeter is operated using a four-key keypad. Two 7-segment LED digital displays show measured values of end-tidal carbon dioxide and saturated oxygen. An LCD graphic display shows the patient’s respiratory rate, pulse rate, and capnograph. The device can be operated in the following three modes: Standard, Homecare, and Sleep Study. The Standard mode is designed for hospital use and intra-hospital transport. The Homecare mode is designed for monitoring in the home by caregivers. The Sleep Study mode includes muted audible and visual queues for use during sleep studies.
The PM100N pulse oximeter includes the name “Nellcor” printed on the front of the device. The model number and other specifications are typically printed either on the top or the side of the device.
Technical Specifications
- Dimensions:
- Width: 10.04"
- Height: 3.23"
- Depth: 6.10"
- Weight: 3 lbs
- Performance
- Measurement Range: SpO2 : 1% to 100%
- Pulse Rate: 20 to 250 beats per minute (bpm)
- Pulse Amplitude: 0.03% to 20%
- Measurement Accuracy
- Saturation
- Adult: 70% to 100% ± 2 digits
- Adult and Neonate Low Sat: 60% to 80% ± 3 digits
- Neonate: 70% to 100% ± 2 digits
- Low Perfusion: 70% to 100% ± 2 digits
- Adult and Neonate with Motion: 70% to 100% ± 4 digits
- Pulse Rate
- Adult and Neonate: 20 to 250 bpm ± 3 digits
- Low Perfusion: 20 to 250 bpm ± 3 digits
- Adult and Neonate with Motion: 20 to 250 bpm ± 5 digits
- Saturation
- Electrical
- Instrument Power requirements: 100 to 240 VAC, 50/60 Hz, 45 VA
- Fuse rating:
- Fast‑acting: 2 A 32VAC/DC
- Fast‑acting 500 mA 32VAC/50DC
- Battery
- Type: Lithium-Ion
- Battery Capacity: Minimum of five hours using new, fully charged battery with no alarms
- Environmental
- Operating Temperature
- Instrument 5 °C to 40 °C (41 °F to 104 °F)
- Transport/Storage Temperature (in shipping carton): -20 °C to 60 °C (-4 °F to 140 °F)
- Operating Humidity: 15% to 93% non condensing
- Operating Altitude: -170 m to 4877 m (-557 ft to 16,000 ft)
- Operating Temperature
- Output
- Trend data download via wired or USB for archiving or data analysis
- Display/Indicators
- Pulse amplitude indicator (eight segments)
- Visual indicators: Pulse search, audible alarms silenced or off, interference indicator, battery charging, and SatSeconds alarm management clock, pleth wave form
- Alarms
- SatSeconds alarm management
- Audible and Visual alarms for high/low saturation and pulse rate, low battery, sensor off, and sensor disconnect
- Categories: Patient status and system status
- Priorities: Low, medium and high
- Notification: Audible and visual
- Setting: Default, institutional and last setting
- Alarm system delay: <10 s
- Available Modes
- Standard – Hospital, hospital‑type facilities, and intra‑hospital transport.
- Homecare – Simplified monitoring for use in the home by caregivers
- Sleep Study – Muted audible and visual queues to aid sleep studies
- Connectivity
- Supports wired and USB trend data export to an external personal computer for archiving or data analysis
- Nurse call capability
Additional Information