Background and Identification

The Model 710 and Model 715 Tidal Wave Sp are handheld, portable capnometers/oximeters made by Respironics, Inc. Pulse oximeters are non-invasive devices used to monitor a patient’s oxygen saturation. Respironics is an American medical supply company that focuses on products designed to improve respiratory functions.

The Model 710 and Model 715 Tidal Wave Sp are designed for monitoring end-tidal carbon dioxide, respiratory rate, functional oxygen saturation, and pulse rate. They can be used in monitoring environments such as emergency, anesthesia, and ventilatory support. The 715 model includes a miniature vacuum pump that draws expired respiratory gases through a CAPNOSTAT carbon dioxide sensor using a nasal cannula and a sampling airway adapter. These devices are designed to monitor adult, pediatric, and neonatal patients.

The 710 and 715 monitors include digital display screens that present the patient’s end-tidal carbon dioxide, respiratory rate, oxygen saturation, and pulse rate. A battery charge indicator and LED light are placed under the left-hand side of the monitor’s display. This LED illuminates when the unit is run on battery power with green indicating a fully-charged battery, slow flashing yellow when battery power is low, and fast flashing red when the battery is exhausted. An LED underneath the monitor’s display provides alert information by presenting either steady yellow light, flashing yellow light, or flashing red and yellow light. A steady yellow light indicates that the audible alarms have been suspended for two minutes or there is no alert in progress. A flashing yellow light indicates that the monitor’s audio has been silenced. A flashing red and yellow light indicates that there is an alert in progress and that the audio has been turned off or suspended.

The name “TIDAL WAVE Sp” is printed in large blue letters across the left-hand side of the monitor’s front panel. The words “NOVAMETRIX CAPNOGRAPHY/OXIMETRY” are also printed in capital letters on the left-hand side of the front face.

Technical Specifications

Capnograph

Principle of Operation: Non-Dispersive İnfrared (NDIR) absorption, dual-wavelength ratiometric- single beam optics

Sensor Type: “Mainstream” (no gas sample drawn from breathing circuit)

Initialization Time: Capnoegram mn 15 seconds, full specifications in 60 seconds.

Response Time: 60 ms

Gas Compensation - Room Air, O2 > 60%, N20 > 60%: Operator selectable in the configuration screen.

Barometric Pressure Compensation: Automatic (range 400-800 mmHg)

CAPNOSTAT™ CO2, Sensor and Airway Adapter:

Weight: Less than 18 g without cable

Sensor Size: 1.3 x 1.67 x .85 inches, (3.30 x 4.24 x 2.16 cm).

Construction: Durable high-performance plastic, ultra-flexible cable

Shock Resistant: CAPNOSTAT™ CO2, sensor withstands a 6-foot drop to a tile floor

Airway Adapter: Single Patient Use, less than 5 cc dead space, meets ANST Z-79

EtCO2 Section (Mainstream)

Range 0-150 mmHg, CO2 partial pressure

Accuracy: 0-40 mmHe +-2 mmHg.41-70 mmHe 5% of readine, 70-150 mmHg +-8% of reading.

Warm-up Time: Operational in 15 seconds, 1 minute to full specifications

Step Response Time: 60 ms, adult: less than 50 ms, neonate

Averaging Time: 1 breath, 10 seconds (default), 20 seconds, instantaneous

Display Resolution: 0-25, 0-50, and 0-150 mmHg in 31 pixels

Alerts: User selectable alert limits for EtCO2

Monitor Specifications:

Classification (IEC6060 1-1): Class 1, internally powered, type BF, enclosure protection rating of IPX1.

Operating Environment: 50 to 104 F (10 to 40° C), 0-90% relative humidity (non- condensing )

Transport/Storage:

  • short term: 14° to 122° F(-10 to 50° C) with NIMH battery
  • long term: 14° to 95° F(-10 to 35° C) with NIMII battery
  • storage: 14° to 131° F(-10° to 55° C) without NIMH battery

Size: 7.97 x 3,25" x 1,5"

Weight: 24 ounces

Power: 100-250 VAC. .38A, 50-60 Hz

Battery: Rechargeable DR30 NiMH battery pack (Cat. No. 400043) or equivalent: AA lithium batteries - Energizer 191 or equivalent.

Battery Life: Approximately 4.5 hours of continuous use with a fully charged DR30 NIMH rechargeable battery pack. Approximately 4.0 hours with sample pump in operation.

Display: LED-backlit 2.5” x 1.25” LCD, adjustable contrast

LED indicators for: Low battery. adapter type. audio/alert status (indicates audio off, 2 minutes suspend, active alert), and external power.