Introduction
For more information on replacing the inspiratory valve, please refer to section 10 in the service manual.
Tools
- 
                  
                  
- 
            
Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.
 
 - 
            
 - 
                  
                  
- 
            
Remove Safety Valve Block
 - 
            
1. Disconnect the cable from the Ambient Valve Solenoid
 - 
            
2. Disconnect the Shaped Hose (A)
 - 
            
3. Remove the 2 Hex Allen Screws (B)
 - 
            
4. Remove the Safety Valve Block
 - 
            
To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block
 
 - 
            
 - 
                  
                  
- 
            
Removing the Inspiration Valve
 - 
            
1. Disconnect the tubing and cables form the Inspiration Valve
 - 
            
2. Disconnect the Shaped Hose (A)
 - 
            
3. Remove 2 Hex Allen screws (B) M5x60
 - 
            
4. Move the Inspiration Valve forward and then up
 - 
            
5. Remove the Inspiration Valve
 
 - 
            
 
To reassemble your device, follow these instructions in reverse order.