Introduction
This guide is for the Philips Respironics V200 / Espirit Ventilator, Part Number 580-1000-02 H. The oxygen water trap/inlet filter assembly consists of a 5-micron (μ) filter to remove particulate (both dry and liquid) from the oxygen gas supply, a bowl with drain for accumulated water, and an oxygen inlet connector (Chapter 3.1.9 on page 3-4 of Service Manual).
Information in this guide was sourced from Chapter 10.1.3 on page 10-5 of the Respironics V200/Esprit Ventilator Service Manual. Further information on general filter replacements for this Ventilator can be found in Chapter 9.1 on pages 9-3 and 9-4 of the Service Manual.
Parts
-
-
Turn off the device.
-
On/Off switch located on the Front Panel as seen in Figure 8-2 (from the Operator Manual).
-
-
-
Disconnect the oxygen hose.
-
Oxygen hose is connected to the Inlet Connector (section 14 of Figure 5-1, diagram from Operator Manual).
-
-
-
Unscrew the water trap bowl (page 10-5 in Service Manual).
-
Located on the back panel of the ventilator (section 15 of Figure 5-1 from Operator Manual).
-
-
-
Reconnect the oxygen hose.
-
According to page 2-2 of the Operators Manual, care in the routing of the oxygen inlet hose should be exercised to ensure it is not exposed to mechanisms that could cause damage by cutting or heating/melting.
-
- Reconnect to AC Power.
- To reassemble your device, follow these instructions in reverse order.