The troubleshooting page for this device can be found here: Respironics V60 Troubleshooting
The Respironics V60 ventilator is an assist ventilator that is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.
Power Supply
- Input: 100 ~ 240 V AC, 50 ~ 60 Hz
- Output: 24 V DC power
- Includes input over-current, output over-voltage, and output current-limiting protection.
Internal Battery
- 14.4 V, 11.5 Ah lithium-ion
- Run-Time: 6 hours (normal conditions)
- Provides operating power when AC power not available
- Internal circuitry monitors battery status, provides self-contained fault control features
General
- Oxygen inlet pressure range: 276 to 600 kPa (40 to 87 psig)
- Weight with optional battery: 10.9 kg (24 lb)
- Weight without optional battery: 10.0 kg (22 lb)
- Dimensions: 33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D)
Patient types
- Suitability: Pediatric (≥ 20 kg) and Adult
Electrical - IntelliBridge EC40/80
- AC voltage: 100 to 240 VAC
- AC frequency: 50 to 60 Hz
- AC power: 300 V·A
- Battery operating time: 6 hours in normal conditions
Settings
- IPAP: 4 to 40 cm H₂O
- C-Flex: Off, 1 to 3
- CPAP: 4 to 25 cm H₂O
- Max P (PPV maximum pressure limit): 5 to 40 cm H₂O
- EPAP: 4 to 25 cm H₂O
- IPAP: 4 to 40 cm H₂O
- Max P (AVAPS maximum IPAP): 6 to 40 cm H₂O
- Inspiratory time: 0.30 to 3.00 sec
- Max P (AVAPS maximum IPAP): 6 to 40 cm H₂O
- Min P (AVAPS minimum IPAP): 5 to 30 cm H₂O
- Min P (AVAPS minimum IPAP): 5 to 30 cm H₂O
- Oxygen percentage: 21 to 100 %
- Ramp time: Off, 5 to 45 min
- Respiratory rate: 4 to 60 beats per minute
- Rise time: 1 to 5
- Triggering and cycling: Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6
- AVAPS target tidal volume: 200 to 2000 ml BTPS
- Max E: 0 to 100 cm H₂O/l
- Max R: 0 to 50 cm H₂Ol/s
- PPV%: 0 to 100 %
- Max P (PPV maximum pressure limit): 5 to 40 cm H₂O
- Max V (PPV maximum volume limit): 200 to 3500 ml
Patient Data
- Breath phase/trigger indicator: Spontaneous, Timed, Exhale
- PIP: 0 to 50 cm H₂O
- Patient/total leak: 0 to 200 l/min BTPS
- Patient Trigger: 0 to 100 %
- Respiratory Rate: 0 to 90 beats per minute
- Ti/Ttot: 0 to 91 %
- Minute volume: 0 to 99.0 l/min BTPS
- Tidal volume: 0 to 3000 ml BTPS
Waveform window
- Pressure waveform: 0 - 50 cm H₂O
- Flow waveform: 240-2401 /min BTPS
- Volume waveform: 0 - 3,500 ml BTPS
Alarm adjustable range
- Lo Rate (low respiratory rate alarm): 1 - 89 beats per minute
- Hi Rate (hi respiratory rate alarm): 5 - 90 beats per minute
- Hi Vr (high tidal volume alarm): 200 - 3,500 ml
- Lo Vr (low tidal volume alarm): OFF - 1,500 ml
- HIP (high inspiratory pressure alarm): 5 - 50 cm H₂O
- LIP (low inspiratory pressure alarm): OFF, 1 - 40 cm H₂O
- LIP T (low inspiratory pressure delay time): 5 - 60 sec
- Lo VE (low minute ventilation alarm): OFF, 0.1 - 991 /min
Credit: Phillips Official Device Page